- Functional
- Powers on. Confirmed by photo. The display lights and reaches a normal working screen (125.0 µl with program P0, Fill and Run indicators and the speed bars lit); every segment renders with no missing, dead or corrupted segments, and the keypad and navigation controls are intact. A Thermo Li-ion pack is installed and a battery symbol IS showing on the display during the test. Pack age is unknown and capacity was not measured. The pipette was powered from a bench DC adapter for this test, so this is a POWERS-ON result; battery runtime was not measured. NOT verified: volumetric accuracy (the calibration sticker has lapsed), a full aspirate/dispense cycle under load, and battery runtime. No runtime figure is claimed. Sold as-is and uncalibrated; budget for a calibration service before analytical use.
- Cosmetic
- Light shop wear consistent with lab use; the LCD is clear and all eight tip cones are present and undamaged. Collar is marked 125 µl. The adjustable tip-spacing rail moves smoothly across its full travel and the scale is legible. A Thermo Li-ion pack is installed and included; the photos show it lifted out of its bay. It is hand-marked with a service date. Third-party calibration stickers from the previous service interval are still on the body.