- Functional
- Powers on. Confirmed by photo. The display lights and reaches a normal working screen (PIPETTE mode, 300 µl, speed setting 8); every segment renders with no missing, dead or corrupted segments, and the keypad and navigation controls are intact. An earlier intake photo also shows the unit lit and prompting PRESS TRIGGER, so the display and firmware both come up normally. No charging cable or cradle is attached to the pipette in the photo. NOT verified: volumetric accuracy (the calibration sticker has lapsed), a full aspirate/dispense cycle under load, and battery runtime. No runtime figure is claimed. Sold as-is and uncalibrated; budget for a calibration service before analytical use.
- Cosmetic
- Light shop wear consistent with lab use; the display window is clear and all twelve tip cones are present, straight and undamaged. Shaft collar is marked 30-300 µl. Handle is marked Thermo Fisher Scientific, Type 14B, Made in Finland. Third-party calibration stickers from the previous service interval are still on the body.